Testimonials

It is very nice to visit Lambda facility and thanks to warm welcome provided by all the teams. I can easily say one of the best CRO I ever visited. People here are wonderful and doing great job

Reputed Client of Israel

Awesome hospitality & fantastic office. All the best!

Reputed Client of Japan

Excellent hospitality and cooperation received throughout the audit. People are knowledgeable and extremely transparent. Looking for long time relation

Reputed Client of Asia

Facility is too good to conduct the BABE studies and adherence to protocol and quality assurance in conducting studies are meeting sponsor expectations and requirements.

Reputed Client of India

Facility is good, people r trained well and know the subject on which they r working. All the best.

Reputed Client of USA

Highly innovative and advanced research centre that possess bright future for the development of the country.

Reputed International Government Body

It is one of the best facility in India for BA/BE

Reputed Regulatory Authority

Lambda is a very professional CRO. All the staff here are experienced and very knowledgeable. I'm grateful for their accommodations and hospitality. The future, if we want to carry out BE study, we will ask Lambda help again.

Reputed Client of Europe

It was a really pleasant meeting with the PV department, and I am looking forward to the next visit.

Reputed Client of Europe

Amazing and best CRO I ever seen

Reputed Client of China

Blog

Blog

Home / Blog

A highly complex, sensitive and simultaneous LC-MS/MS method for the quantification of Total Estrone & Total Equilin in human plasma

21 Aug 2023

Estrone is one of the major naturally occurring (endogenous) steroidal estrogens, while Equilin is classified as a non-endogenous estrogen. They are used for the management of perimenopausal and postmenopausal symptoms.

Study Design

Equilin
Estrone

Since both molecules have their in vivo metabolites, namely sulfate and glucuronide, there is a regulatory requirement to establish a bioanalytical method for assessing both the free and conjugated components (i.e., total Estrone and total Equilin) in clinical study samples. At Lambda Therapeutic Research, we have developed and validated a highly sensitive bioanalytical method capable of quantifying total Estrone and total Equilin in a single analytical run.

Challenges

Developing a method for quantifying low-dosage, endogenous Estrone and non-endogenous Equilin required innovative approaches to handle interference-free matrix preparation, efficient enzymatic conversion, and matrix effect mitigation, while also addressing derivatization needs and potential interference from structurally similar compounds in the biological matrix.

  • Both molecules belong to the families of steroids/hormones, and their dosages are extremely low, resulting in detection levels in the low picogram range. Since Estrone is an endogenous compound, the need for a surrogate matrix to create an interference-free matrix for the preparation of CC/QC samples arises. Hence, a charcoal stripping method was developed to prepare interference-free human plasma. On the other hand, Equilin is a non-endogenous compound, and developing a single-run method for Equilin posed challenges in minimizing matrix effects between the surrogate and actual matrix.
  • Due to the requirement for developing a method for total Estrone and Equilin (as conjugated forms), it was necessary to convert sulfate and glucuronide metabolites into their respective parent molecules. To achieve this, three different types of enzymes were selected and evaluated. One enzyme was chosen that exhibited efficient sulfatase and glucuronidase activity in the actual matrix. A unique set of conditions involving incubation time, temperature, and buffer solution addition for sample preparation was optimized to ensure proper conversion and accuracy.
  • Given the widely accepted practice to ionize steroids/hormones in triple quadrupole mass spectrometry, derivatization becomes essential for achieving low picogram levels in a biological matrix. Hence, after enzymatic reaction and sample cleanup using solid-phase extraction, a derivatization reaction was implemented prior to LC-MS/MS analysis.
  • Due to the presence of structurally similar compounds and related hormones in the formulation and biological matrix, the potential for interference with targeted analyte chromatography was significant. To address this challenge, we employed a unique core-shell technology-based analytical column.

At Lambda, we have developed a multi-step and robust method encompassing enzymatic reaction, solid-phase extraction, derivatization, and liquid-liquid extraction. This comprehensive approach ensures both acceptable precision and accuracy while effectively addressing matrix effect-related concerns.

Extraction Method flow


Chromatography of LLOQ sample

Estrone
Equilin

Results

Below are the global statistics for three precision and accuracy runs conducted on both altered and unaltered matrices.

Equilin - Global Statistics (Altered matrix QCs)
HQC MQC LQC LOQQC
Mean (pg/mL) 7791.4368 4357.4172 118.5582 40.4250
SD± 183.87361 111.02256 3.13429 3.23238
Precision (%CV) 2.4 2.5 2.6 8.0
Nominal value(pg/mL) 7952.346 4373.791 119.711 41.500
Accuracy % 98.0 99.6 99.0 97.4
n 18 18 18 18

Estrone - Global Statistics (Altered matrix QCs)
HQC MQC LQC LOQQC
Mean (pg/mL) 7564.2454 4315.3342 116.2711 43.7111
SD± 159.65513 102.64951 2.89671 2.34112
Precision (%CV) 2.1 2.4 2.5 5.4
Nominal value(pg/mL) 7950.993 4373.046 119.690 41.493
Accuracy % 95.1 98.7 97.1 105.3
n 18 18 18 18

Equilin - Global Statistics (Un-altered matrix QCs)
HQC MQC LMQC LQC LOQQC
Mean (pg/mL) 7059.6958 3949.4333 1009.5493 116.2473 35.6806
SD± 187.70029 109.59439 35.68931/td> 4.60135 3.14243
Precision (%CV) 2.7 2.8 3.5 4.0 8.8
Nominal value (pg/mL) 7930.991 4481.009 1053.037 118.795 40.391
Accuracy % 89.0 88.1 95.9 97.9 88.3
n 18 18 18 18 18

Estrone - Global Statistics (Un-altered matrix QCs)
HQC MQC LMQC LQC LOQQC
Mean (pg/mL) 7383.5671 4180.9837 1145.2358 107.2721 47.1621
SD± 270.78552 141.94408 53.68607 10.60584 4.85764
Precision (%CV) 3.7 3.4 4.7 9.9 10.3
Nominal value (pg/mL) 8012.494 4580.041 1169.486 107.697 44.662
Accuracy % 92.2 91.3 97.9 99.6 105.6
n 18 18 18 18 18

We have generated matrix stability data, including benchtop stability for 13.0 hours, freeze-thaw stability for five cycles, and long-term stability in matrix for 81 days as part of the method validation process.

Application

The validated method has been successfully applied to analyze study samples in two different studies, in which incurred sample reproducibility experiment was conducted. The results were found to be well within the acceptable criteria for both analytes, thus demonstrating the method's reproducibility for incurred samples.

Conclusion

A selective, precise, accurate, simultaneous, and highly sensitive LC-MS/MS method is now available for the quantification of Total Estrone and Total Equilin in human plasma.